IDEAL臭氧在线检测仪英文说明书.docx
《IDEAL臭氧在线检测仪英文说明书.docx》由会员分享,可在线阅读,更多相关《IDEAL臭氧在线检测仪英文说明书.docx(11页珍藏版)》请在冰豆网上搜索。
![IDEAL臭氧在线检测仪英文说明书.docx](https://file1.bdocx.com/fileroot1/2023-1/22/734e497d-0c67-45ef-9f47-14660a6769ed/734e497d-0c67-45ef-9f47-14660a6769ed1.gif)
IDEAL臭氧在线检测仪英文说明书
IDEAL-2000
Ozoneconcentrationdetector
Manual
UnitedStates-ownedIDEALComputer(ZIBO)CO,.LTD.
Dearusers:
Thankyouverymuchforusingourproducts.
Beforeusingthisproduct,pleasereadtheseinstructionscarefully.
Theproductsdescribedinthismanualonlyfortrainedqualifiedworkers.
Thisproduct'sadjustment,repairandmaintenancemustbeauthorizedofficersspecifiedbythemanufacturer.
Forusethisproductcorrectlyandsafety,youneedtooperateinaccordancewiththismanual.
Themanufacturerrecommendations:
Maintaintheintegrityoftheproductlabel;
Donotdisassemblethisproduct;
AllmaintenanceandcalibrationofthisproductneedstooperateinaccordancewiththerequirementsoftheInstructionManual.
Themanufacturerhasnoresponsibilitytoanydamageandinjurecausedbyimproperuseofthisproduct.
Themanufacturercanimprovetheproductspecificationsormodifythecontentsofthemanualwithoutnotice.
Ifyoufindanyerrorsinthismanual,pleasecontactus,weareverygrateful.
DONOTcopyorimitationofallorpartofthismanualwithoutThemanufacturer’sPermission!
Directory
1.Outline
2.TechnicalParameters
3.Structure
4.FunctionIntroduction
5.Communicationprotocol
6.Maintenance
7.Notes
8.Store
9.dateofvalidity
1.Outline
IDEAL-2000-typeOzoneconcentrationdetectormainlyusefortheexport’sozoneconcentrationdetectionofozonegenerator,UVlampaslightsourcesystem,Lifeis30,000hoursormore,Stablemeasurements,Highprecision;UseIntegrationLayerStreamlightpoolstructure,Noleakage、absorbhighpressure、absorbhighflowsamplegas’simpact、easytoclean、easytomaintenance、easytooperate.Theproductwillcorrectionzeroautomaticallyduringusing,ensureaccuracyofconcentrationandavoidzerodrift,have4-20mAandRS-485outputsignal。
IDEAL-2000-typeOzoneconcentrationdetectorisapplicationofthephotoelectricconversiondetectiontechnology,usethemicroprocessorcontroltechnology,longlifepulsemodelightsource,usethecharacteristicsofUVabsorptionofozone,accordancewithTheBeer-LambertLaw.
2.TechnicalParametersandInstallation
1.TechnicalParameters:
Measuringrange:
0~200mg/L;
Displayresolution:
0.1mg/L;
Displaydigits:
4;
marginoferror:
±3%;
Communicationmethod:
RS-485;
CommunicationParameters:
9600,8,N,1;
Outputmethod:
differential4-20mA,Linear;
Weight:
1.5kg;
Inputgasflow:
1.0±0.3L/min;
InputgasPressure:
MAX1bar(0.1MPa);
Powersupply:
AC220V±10%50Hz;
Size:
160mm(W)×80mm(H)×250mm(L);
Gasinputport、outputportID4mm,OD6mm;
Embeddedinstrumentationsize:
152mm×76mm;
Cycleofbacktomanufacturertocheck:
1.5years.
IDEAL-2000Ozoneconcentrationdetector
2.TheamendcalculationoftheconcentrationofOzoneGeneratoroutletinsidepipeline
IDEAL-2000Ozoneconcentrationdetector`sdetectionoftheozoneconcentrationisusuallytheconcentrationofozoneatatmosphericpressure(andthecorrespondingstandardiodometricmethod),ifthecustomerwouldliketomeasuretheozonegeneratorexportpipelinepressureofozoneintheozoneconcentrationundertheconditions,calculatedaccordingtothefollowingformula:
(PNTP+P1)×(TNTP+Ta)
Camend=Ca×————————————
(PNTP+Pa)×(TNTP+T1)
PNTP:
Standardatmosphere101.325KPa
Pa:
PressureinsidethepipeoftheOzoneconcentrationdetector(Generally0KPa)
P1:
thepressureofthemainexportpipeoftheOzoneGenerator(KPa)
Ca:
ReadingoftheOzoneconcentrationdetector(mg/L)
Ta:
TheinsidetemperatureofthepipeofOzoneconcentrationdetector(℃)
T1:
TheinsidetemperatureofthemainexportpipeoftheOzoneGenerator(℃)
TNTP:
Standard0℃=273.15K(K)
NOTE:
GenerallycircumstancesthetemperaturevaluesTaandT1isthesame
3.Structure
ThefrontpanelandthebackpanelofIDEAL-2000-typeOzoneconcentrationdetectorasshown:
`
Theschematicdiagram
1Thefrontpanel2Screen
3Zerocalibrationbutton4Thebackpanel
5Powerswitch6Ozoneinput
7Ozoneoutput8Airconnector
9Poweroutlet10Nameplate
11StatusIndicator(greenisdetecting,rediszerocalibrating)
12Ozonereducingagentsealing
Airconnectorpins:
1.RS-485TX+[Brown]2.RS-485TX-[Red]
3.Currentoutput(4-20mA)CO+[Orange]4.Currentoutput(4-20mA)CO-[Yellow]
Gasinputport、outputportID4mm,OD6mm。
4.FunctionIntroduction
Thedetectorisswitchedintotheworkingstate.After30minutesofwarm-useproposals.
Whenusing,pleasefirstensurethattheelectrical,pipingconnectedcorrectly,especiallyozoneinputandoutputportscannotbereversed,flowcontrolaroundthe1.0L/min.
ThedetectorisautomaticZerocalibrationwhenboot,toRestoretheinputgastozeroair.Automaticcalibrationintervalisgradualextensionfrom3minutes,afterabout30minutes,itwillbe10minutesonce.UserscanalsosendcommunicationsthroughacommandorbuttontocalibrationZero.
Useofthisdetector:
1electrical,pipingproperlyconnected;
2Poweron,preheating30minutes;
3Ozoneconcentrationmeasurement;
4Data,output,processinganddatacommunicationbuttonoperation;
5Whenmeasurement,iftheerrorisgreaterthanonepercent,itwillautomaticallycalibrationzero.
6whenreadingorwritingRS-485address,ensurethatonlyonebusfromthemachine,otherwise,addressconfliction.
NOTE:
1、Detectionofgasdewpointshouldbelessthan5℃,Otherwise,gasshouldbedried.
2、Detectormustoperateintheabsenceofcondensingenvironment.
3、Monitorrun-timetoensurestablegaspathflow,flow1.0L/min
4、Testingprocess,shouldberecycledordecompositionofozonegas,ozonegasmaynotleaktotheindoorair.
5、Prohibitairconnector3,4(currentsignaloutput)ofanysidegrounded.
5.Communicationprotocol
1、Outline
The"OzoneMonitorcommunicationprotocol",describingtheozonedetectorwiththeoutsideworldviaserialcommunicationportfordataexchangeformatandcontentofthedataforparametersettingfunctionoftheozonedetectortestanddatamonitoring.TheagreementwasfinalizedversionV1.0.
2、ProtocolDescription
2.1Physicalinterface
SerialcommunicationinterfaceRS-485。
Serialportparametersis“9600,n,8,1”。
2.2CommunicateMethod
Answermode。
Allactionsarelaunchfromoutsidesoftware.
2.3Basicformat
CD
SNB
CommandNote
SOI
1
Startbit
LENGTH
2
Datalength
ADD1
3
Communicationinitiatoraddress
ADD2
4
Recipientaddresscommunication
CID
5
Commandidentifier
DATAINFO
6toLENGTH+1
Data
CHKSUM
LENGTH+2
CheckCode
EOI
LENGTH+3
Endbit
2.4Formatexplanation
●Thedataintheprotocol,allare16-bandformat,highintheformer.InadditiontoSOIandEOIisahexadecimalinterpretation(SOI=16H,EOI=17H),hexadecimaltransmission,therestoftheexplanationareinhexadecimal,hexadecimal-ASCIIcodeofthewaytransmission,eachbyteofthetwoASCIIcodes,ietheASCIIcodeofahighfour,saidwithanASCIIcodelowerfoursaid.Cases:
CID=26H,theorderoftransmissionand36H32Hsenttwobytes.
●SOI:
StartofInformation,Framestartcharacter,orinitiation;a16-bytehexadecimalnumber,16H;
●LENGTH:
LENGTHlater(notincludingLENGTH)tocheckcodedatalength(16hexadecimalnumber);a16-bytehexadecimalnumber;
●ADD1:
Communicationinitiatoraddress,hostaddressis00H~7FH;
●ADD2:
Communicationsrecipientaddress,theslaveaddressis80~0FFH;
●CID:
CommandIDCommandidentifier,commanddescribedtheoperationofactorbreakdown;a16-bytehexadecimalnumber;
●DATAINFO:
Commandordatainformation;(LENGTH-4)bytesof16hexadecimalnumbers,thatis,LENGTH=4didnotcommandordatainformation;
●CHKSUM:
Framecheckcode;a16-bytehexadecimalnumber
●Foreachinstruction,thereceiverneedstodothefollowingcheck:
firstcheckthelengthcheck,framecheck;
●CHKSUMcalculation:
startingfromtheLENGTH(includingLENGTH),toDATAINFO(includingDATAINFO),all16hexadecimalnumbersXORlogic;
●EOI:
EndofinformationFrameterminator,orendofthebit;a16-bytehexadecimalnumber;17H.
3、ProtocolOverview
InaccordancewiththeoperationofthesoftwareoutsidethepurposeofthisAgreementtocompletetheoperationinstructionisdividedintothefollowingcategories,theagreementusedtodistinguishthetypeofoperationCID:
⏹setparameterscommand:
Tomodifytheparametersoftheozonedetector.
⏹readbackparameterscommand:
toreadbacktheparametersoftheozonedetector,uploadOzoneMonitortomonitortheozoneconcentration.
⏹commandparametercalculation:
Inordertorecalculatetheparametersoftheozonedetector.
4、Protocol,withoutSOI、LENGTH、CHKSUMandEOI
CB
Function
00H
ReadRS-485address
0FFH
WriteRS-485address
20H
Reserved
21H
Reserved
23H
Reserved
24H
Reserved
25H
Reserved
2FH
Readozoneconcentration
30H
Reserved
33H
Reserved
34H
Reserved
35H
Reserved
3DH
Writeconcentrationofzerocorrection
3EH
Writeconcentrationoffactorcorrection
40H
Reserved
Reservedcommandwords,donotoperate,ormaymakeequipmentworkexception.
Setcommandsuccessfulresponsecommandwordthesame,theresponsedataforthe0CCH,setthecommandfailedtorespondtocommandsforthe66H,sendthecommandresponsedatafortheword.
4.1Readingbackthecommandparameters
⏹ReadingOzoneDetectorRS-485address
Senddataformat:
CID
DATAINFO
00H
DATAINFOisdata,thisorderNodata
Responsedataformat:
CID
DATAINFO
00H
DATAINFOisreturndata,thatisRS-485address,1byte,unsignedtypeofdata.
Readingozoneconcentration
Senddataformat:
CID
DATAINFO
2FH
DATAINFOisdata,thisorderNodata
Responsedataformat
CID
DATAINFO
2FH
DATAINFOisReturndata,thatisozoneconcentration,2bytes,signed-typedat