1、IDEAL臭氧在线检测仪英文说明书IDEAL-2000Ozone concentration detectorManualUnited States-owned IDEAL Computer(ZIBO) CO,.LTD.Dear users:Thank you very much for using our products.Before using this product, please read these instructions carefully.The products described in this manual only for trained qualified wor
2、kers.This products adjustment, repair and maintenance must be authorized officers specified by the manufacturer.For use this product correctly and safety, you need to operate in accordance with this manual.The manufacturer recommendations:Maintain the integrity of the product label; Do not disassemb
3、le this product; All maintenance and calibration of this product needs to operate in accordance with the requirements of the Instruction Manual.The manufacturer has no responsibility to any damage and injure caused by improper use of this product. The manufacturer can improve the product specificati
4、ons or modify the contents of the manual without notice.If you find any errors in this manual, please contact us, we are very grateful.DO NOT copy or imitation of all or part of this manual without The manufacturers Permission!Directory1. Outline2. Technical Parameters3. Structure4. Function Introdu
5、ction5. Communication protocol6. Maintenance7. Notes8. Store9. date of validity1. OutlineIDEAL-2000-type Ozone concentration detector mainly use for the exports ozone concentration detection of ozone generator,UV lamp as light source system,Life is 30,000 hours or more,Stable measurements, High prec
6、ision;Use Integration Layer Stream light pool structure,No leakage、absorb high pressure、absorb high flow sample gass impact、easy to clean、easy to maintenance、easy to operate. The product will correction zero automatically during using, ensure accuracy of concentration and avoid zero drift,have 4-20m
7、A and RS-485 output signal。IDEAL-2000-type Ozone concentration detector is application of the photoelectric conversion detection technology, use the microprocessor control technology,long life pulse mode light source,use the characteristics of UV absorption of ozone,accordance with The Beer-Lambert
8、Law.2. Technical Parameters and Installation1Technical Parameters:Measuring range:0200mg/L;Display resolution:0.1mg/L;Display digits:4;margin of error:3;Communication method:RS485;Communication Parameters:9600,8,N,1;Output method:differential 4-20mA,Linear;Weight:1.5kg;Input gas flow:1.00.3L/min;Inp
9、ut gas Pressure:MAX 1bar(0.1MPa);Power supply:AC 220V10% 50Hz;Size:160mm(W)80mm(H)250mm(L);Gas input port、output port ID 4mm,OD 6mm;Embedded instrumentation size:152mm76mm;Cycle of back to manufacturer to check:1.5 years.IDEAL-2000Ozone concentration detector 2The amend calculation of the concentrat
10、ion of Ozone Generator outlet inside pipelineIDEAL-2000 Ozone concentration detectors detection of the ozone concentration is usually the concentration of ozone at atmospheric pressure (and the corresponding standard iodometric method), if the customer would like to measure the ozone generator expor
11、t pipeline pressure of ozone in the ozone concentration under the conditions, calculated according to the following formula: (PNTP+P1) (TNTP+Ta)Camend=Ca (PNTP+Pa) (TNTP+T1)PNTP:Standard atmosphere 101.325KPaPa:Pressure inside the pipe of the Ozone concentration detector(Generally 0KPa)P1:the pressu
12、re of the main export pipe of the Ozone Generator(KPa)Ca:Reading of the Ozone concentration detector(mg/L) Ta:The inside temperature of the pipe of Ozone concentration detector()T1:The inside temperature of the main export pipe of the Ozone Generator()TNTP:Standard 0 =273.15K(K)NOTE:Generally circum
13、stances the temperature values Ta and T1 is the same3. StructureThe front panel and the back panel of IDEAL-2000-type Ozone concentration detector as shown:The schematic diagram 1 The front panel 2 Screen3 Zero calibration button 4 The back panel5 Power switch 6 Ozone input7 Ozone output 8 Air conne
14、ctor9 Power outlet 10 Nameplate11 Status Indicator(green is detecting,red is zero calibrating)12 Ozone reducing agent sealingAir connector pins:1. RS-485 TX+ Brown 2. RS-485 TX- Red3. Current output (4-20mA) CO+ Orange 4. Current output (4-20mA) CO- YellowGas input port、output port ID 4mm,OD 6mm。4.
15、Function IntroductionThe detector is switched into the working state. After 30 minutes of warm-use proposals.When using, please first ensure that the electrical, piping connected correctly, especially ozone input and output ports can not be reversed, flow control around the 1.0L/min.The detector is
16、automatic Zero calibration when boot, to Restore the input gas to zero air. Automatic calibration interval is gradual extension from 3 minutes, after about 30 minutes, it will be 10 minutes once. Users can also send communications through a command or button to calibration Zero.Use of this detector:
17、1 electrical, piping properly connected;2 Power on , preheating 30 minutes;3 Ozone concentration measurement;4 Data, output, processing and data communication button operation;5 When measurement, if the error is greater than one percent, it will automatically calibration zero.6 when reading or writi
18、ng RS-485 address, ensure that only one bus from the machine, otherwise, address confliction.NOTE:1、Detection of gas dew point should be less than 5,Otherwise, gas should be dried.2、Detector must operate in the absence of condensing environment.3、Monitor run-time to ensure stable gas path flow, flow
19、 1.0L/min4、Testing process, should be recycled or decomposition of ozone gas, ozone gas may not leak to the indoor air.5、Prohibit air connector 3,4 (current signal output) of any side grounded.5. Communication protocol1、OutlineThe Ozone Monitor communication protocol, describing the ozone detector w
20、ith the outside world via serial communication port for data exchange format and content of the data for parameter setting function of the ozone detector test and data monitoring. The agreement was finalized version V1.0.2、Protocol Description2.1 Physical interfaceSerial communication interface RS-4
21、85。Serial port parameters is“9600,n,8,1”。2.2 Communicate MethodAnswer mode。All actions are launch from outside software.2.3 Basic formatCDSNBCommand NoteSOI1Start bitLENGTH2Data lengthADD13Communication initiator addressADD24Recipient address communicationCID5Command identifierDATAINFO6 to LENGTH+1D
22、ataCHKSUMLENGTH +2Check CodeEOILENGTH +3End bit2.4 Format explanation The data in the protocol, all are 16-band format, high in the former. In addition to SOI and EOI is a hexadecimal interpretation (SOI = 16H, EOI = 17H), hexadecimal transmission, the rest of the explanation are in hexadecimal, hex
23、adecimal-ASCII code of the way transmission, each byte of the two ASCII codes, ie the ASCII code of a high four, said with an ASCII code lower four said. Cases: CID = 26H, the order of transmission and 36H 32H sent two bytes. SOI:Start of Information,Frame start character, or initiation; a 16-byte h
24、exadecimal number, 16H; LENGTH:LENGTH later (not including LENGTH) to check code data length (16 hexadecimal number); a 16-byte hexadecimal number; ADD1: Communication initiator address, host address is 00H7FH; ADD2: Communications recipient address, the slave address is 800FFH; CID:Command ID Comma
25、nd identifier, command described the operation of act or breakdown; a 16-byte hexadecimal number; DATAINFO:Command or data information; (LENGTH - 4) bytes of 16 hexadecimal numbers, that is, LENGTH = 4 did not command or data information; CHKSUM:Frame check code; a 16-byte hexadecimal number For eac
26、h instruction, the receiver needs to do the following check: first check the length check, frame check; CHKSUM calculation: starting from the LENGTH (including LENGTH), to DATAINFO (including DATAINFO), all 16 hexadecimal numbers XOR logic; EOI:End of information Frame terminator, or end of the bit;
27、 a 16-byte hexadecimal number; 17H.3、Protocol OverviewIn accordance with the operation of the software outside the purpose of this Agreement to complete the operation instruction is divided into the following categories, the agreement used to distinguish the type of operation CID: set parameters com
28、mand: To modify the parameters of the ozone detector. read back parameters command: to read back the parameters of the ozone detector, upload Ozone Monitor to monitor the ozone concentration. command parameter calculation: In order to recalculate the parameters of the ozone detector.4、Protocol,witho
29、ut SOI、LENGTH、CHKSUM and EOICBFunction00HRead RS-485 address0FFHWrite RS-485 address20HReserved21HReserved23HReserved24HReserved25HReserved2FHRead ozone concentration30HReserved33HReserved34HReserved35HReserved3DHWrite concentration of zero correction3EHWrite concentration of factor correction40HRes
30、ervedReserved command words, do not operate, or may make equipment work exception.Set command successful response command word the same, the response data for the 0CCH, set the command failed to respond to commands for the 66H, send the command response data for the word.4.1 Reading back the command
31、 parameters Reading Ozone Detector RS-485 addressSend data format:CIDDATAINFO00H DATAINFO is data, this order No dataResponse data format:CIDDATAINFO00HDATAINFO is return data, that is RS-485 address, 1 byte, unsigned type of data.Reading ozone concentrationSend data format:CIDDATAINFO2FH DATAINFO is data, this order No dataResponse data formatCIDDATAINFO2FH DATAINFO is Return data, that is ozone concentration, 2 bytes, signed-type dat