1、中性压力记录仪使用说明书英文版CONTENTSChapter 1 Summarize . 2Chapter 2 Main features 2Chapter 3 Performance index . .3Chapter 4 Contour dimension 4Chapter 5 Installation .51、Direct-connected .62、Trestle installation .7Chapter 6 Keypad instructions 8Chapter 7 Operation .91、Time setting . 92、Interrecord-gap setting
2、.113、The number of storing data query134、Data file name Settings 135、Zero correction 146、Full-range correction 147、Start/stop recording .158、Read data. .169、Instrument calibration. .1710、Battery replacement. .18Chapter 8 Management software using 19Chapter 9 Failure analysis and ruled out 26Chapter
3、10 Maintenance and transportation26 1、Maintenance. .262、Transportation and storage26Chapter 1 Summarize Pressure Recorder is a kind of electronic data record device which specially detects the pressure of a pipe or a pressure vessel .It sets the sensor, intelligent instrument at an organic whole, ca
4、n collect and record pressure data according to the setted pressure record time interval. Pressure data can display through the liquid crystal , also can be read to the computer through USB, and cooperate with PC management software to statistics and analyse the data.Chapter 2 Main features Paperles
5、s 、Low power consumption Small in size、 High integrated、For easy installation Liquid crystal display measured pressure value and historical data on site Time interval of pressure record: 1 second-18 hours、Can input configuration via keyboard Record the most up to 10 years of history pressure data Bu
6、ilt-in high-capacity storage to store data , Can store 20000 data Read data via U-disk Chapter 3 Performance index Ambient temperature:-2060 Ambient humidity:10%95% Process fluid measured:noncorrosive gas、liquid Precision:0.5 degree Grade of waterproof:IP65 Grade of explosive-proof:ExibBT4Chapter 4
7、Contour dimension Unit:mmChapter 5 InstallationPressure recorder is linked with joining pipe through the thread interface, the connection need to install needle valve to be convenient to debugging, replacement and maintenance.Measuring installation points should as far as possible to avoid the inter
8、ference of environmental factors, such as severe temperature changing, vibration and impact etc.The thread interface of Pressure recorder is M20 x 1.5, currently two installations are recommended: the Pressure recorder directly linked with the needle valve ; the Pressure recorder installed in the wa
9、ll with stents (or tubes of curved stents or tubes of flat stents), and then connects to needle valve.It is shown as Figure 5.1 below:Figure 5.11、Direct-connected When it is Direct connection, needle valve interfaces are allowed to be linked with two types of threaded interface through Adapter, metr
10、ic thread M20 x 1.5 and 1/2 NPT thread. Installation procedures are as follows:1) When the meter needle valve interface is M20 x 1.5 Install the needle valve in the pressure tap to make sure that the joint takes good seal and needle valve is in the closed position; Install Pressure recorder, the int
11、erface of Pressure recorder and the interface of the stop valve are sealed via seal gasket; Open the stop valve, observe the indicating value of Pressure recorder and set the Pressure recorder.2) When the meter needle valve interface is 1/2 NPT Select NPT thread adapter (optional), install it in the
12、 interface of Pressure recorder; Install the needle valve in the pressure tap to make sure that the joint takes good seal and needle valve is in the closed position; Install Pressure recorder,wind the adhesive tape between the interface of Pressure recorder and the meter needle valve interface Open
13、the needle valve, observe the indicating value of Pressure recorder and set the Pressure recorder.2、Trestle installation Confirm the installation position of Pressure recorder, fasten Pressure recorder; Confirm catheter trend and length according to the distance between installation position and Pre
14、ssure tapping point , and site equipment distribution; Check whether pressure tapping point takes good sealing; Choose stainless steel pipe whose diameter is 6 as joining pipe to bend. Connect joining pipe with the interface of Pressure recorder s guardstaple; Figure 5.2 Check the sealing, open need
15、le valve, observe the indicating value of Pressure recorder and make the function setting. It is shown as Figure 5.2.Chapter 6 Keypad instructions1 Calibration key: Calibrate instrument when the precision is incorrect2 Data read key:Read pressure data with U-disk.3 Start key: start/stop recording pr
16、essure data4 Up key:Used to scroll the numerical values up.5 Setting key:Used to set the parameters 6 Down key:Used to scroll the numerical values down7 Reset key:Troubleshooting, data resetChapter 7 Operation1、Time settingThe pressure data recorded include time information,this time information is
17、from the internal clock unit of the instrument,after long time running this clock unit will have error,regular adjustment is necessary,to the users who require exact time it is necessary to adjust once two months.It must be adjusted after replacing the battery or power off. The method is as follows:
18、Press “”for half a second,display “1t1”,loosen the buttonUntil set in the required valuesPress“”to increase,press “”to decrease Press “”for half a second,access the setting of “Year” Press “”for half a second,access the setting of “Month”Similarly set “Day”“Hour” “Minute”“Second”in the required valu
19、es,display“EC”,and loosen the button,exit setup.Until set in the required valuesPress“”to increase,press “”to decrease Menus and symbols instructions: Displayed symbol Meaning enter to set yaar setting month setting day setting hour setting minute setting second setting exit setup 2、Recording interv
20、al settingThis instrument has the function of recording pressure data, recording mode is constantduration recording,the interval can be adjusted between 1 second to 18 hours.The method is as follows:Until set in the required valuesPress“”to increase,press “”to decrease Press “”for half a second,acce
21、ss the setting of “Hour”Press “”for half a second,display “1t1”,loosen the button .Press“”for half a second,display “2 In”loosen the button. Press “”for half a second,access the setting of “Minute”Similarly set “Second”in the required value,display“EC”,and loosen the button,exit setup.Until set in t
22、he required valuesPress“”to increase,press “”to decrease Menus and symbols instructions:Displayed symbol Meaning enter to set recording interval setting hour setting minute setting second setting exut setup3、The number of storing data queryPress for half a second, show “1t1” , and loosen the key. Pr
23、ess for half a second, show2 In, and loosen the button. Press for half a second, show 3 AC, loosen the button. Press for half a second, the instrument shows the following menu: , for example: instrument shows r 0050,indicate that the recorder has recorded 50 data.Press for half a second, showEC, and
24、 loosen the key, instrument exit number query, the recorder switch to the screen of normal pressure value automatically.4、 Data file name SettingPress for half a second, show“1t1”, and loosen the button. Press for half a second, show 2 In, and loosen the button. Press for half a second, show 3 AC, l
25、oosen the button. Click“”again, show 4 SA . Pressfor half a second, enter the interface of setup, show scene as follows, press to increase, press to decrease until to the required name. Press for half a second, showEC, and loosen the key, the instrument exit from set condition, data file name settin
26、g completed.Note: the data file name can only be four digits.5 、Zero correctionPress for half a second, show “1t1”, and loosen the button. Press until the screen shows “5 OP”. Press , show OP 0.0000, the 0.000 is zero, you can adjust the number via the direction key , press until the accurate number
27、, correction completed. And then into the measurement model, at this time the pressure value showed has been corrected.6 、Full-range correctionFull-range correction not always made in the largest of the range,it can be corrected as long as the pressure is more than half range and less than or equal
28、to full range, but the more close to the full range the better the correction result. When the value is less than half the range the correction operation is forbidden.During correction the pressure shoud be ensured stable and no changing.Press to enter the menu, press until the 6SP showed. And then
29、press to enter full range correction. At this time the screen shows SP00.00. The 00.00 is the pressure value before you entering the menu. Press to enter the menu, press until the 6SP showed. And then press to enter full range correction. At this time the screen shows SP00.00. The 00.00 is the pressure value before you entering the menu.The value may be different from the ideal value, you can adjust the value via the direction key , press until the accurate number, correction completed. And then into the measurement