ConductivityMeasuringCells

forHigh-TemperatureApplicationsConduMaxWCLS12

Two-electrodemeasuringsensorswithcellconstantk=0.01/cmork=0.1/cm

TechnicalInformationTI082C/24/ae

Areasofapplication

TheconductivitymeasuringcellCLS12isdesignedforindustrialandpowerplantapplications(e.g.condensatemeasure-mentwherelowconductivitiesmustbemeasuredunderhighpressuresandtemperatures.

Themeasuringrangeofthecellsdependsonthecellconstantk.

•k=0.01/cm:

0.04to20µS/cm•k=0.1/cm:

0.1to200µS/cm

MeasuringcellswithaPt100temperaturesensorareusedtogetherwithconductivitymeasuringinstrumentsequippedwithautomatictemperaturecompensation:

•MycomSCLM153

•LiquisysMCLM223/253•MyProCLM431

Formeasurementofspecificresistance,MΩ·cmmeasuringrangesareavailableinthemeasuringtransmittersoftheMycomandMyProseries.

Benefitsataglance

•Optimumadaptationtoprocess

conditionsormountingplaceduetodifferentdesigns

•Mountinginpipesorflowchambers•Applicationwithtemperaturesofupto320°F(160°Candpressuresofupto580psi

•Simplemeasuringcableconnectionduetolargeconnectioncompartment

•Qualitycertificatestatingtheindividual

cellconstant

ThePowerofKnowHow

ConduMaxWCLS12

Endress+Hauser

Functionandsystemdesign

Conductiveconductivitymeasurement

Theconductivityofliquidsismeasuredwithameasuringsystemthathastwocoaxiallyarrangedelectrodessimilartoacapacitor.Theelectricresistanceoritsreciprocalvalue,theconductanceG,ismeasuredaccordingtoOhm’slaw.ThespecificconductivityKisdeterminedusingthecellconstantkthatisdepen-dentonthesensorgeometry.

Measuringprinciple

ConductiveconductivitymeasurementAC=PowersupplyI=CurrentmeterV=Voltagemeter

ConduMaxWCLS12importantproperties

•Electrodes

ConduMaxWCLS12hasespeciallylargecoaxialmeasuringelectrodes.Theseallowhighflowratesandhighaccuracy•Temperaturecompensation

Inaddition,aPt100temperaturesensorisinstalledintheinsideelectrodetomeasuretheprocesstemperature.

Thehighconductivityallowsfastadjustmentofthesensortotheprocess

temperature,therebyguaranteeingpreciseautomatictemperaturecompensationintheconnectedmeasuringtransmitter.•Durability

Thesensorsaredistinguishedbyhighthermal,chemicalandmechanicalresistances.Themaximumoperatingpressureis580psi(40barupto212°F(100°C,themaximumoperatingtemperatureis320°F(160°Cforupto87psi(6bar.

•Useinsuperheatedsteamorultrapurewater

Specialsealingmaterialsareavailableforuseinsuperheatedsteamorultrapurewaterathightemperatures.Fortheseapplications,nonominaldurabilityofthestandardEPDMmaterialcanbegiven.

2

ConduMaxWCLS12

Endress+Hauser3

Measuringsystem

Acompletemeasuringsystemconsistsof:

•ACLS12conductivitysensor

•Atransmitter,e.g.MycomSCLM153

•ACYJK71orCYK71-Exmeasuringcable

ACLA751flowchamber(refertoaccessoriesisavailbe(optionallywhichcanbeusedforinstallationintotheprocess.

Measuringsystemexample1

CLS12ConduMaxWsensor2CLM153MycomStransmitter3Measuringcable

Input

ConductivityandtemperatureDependentonversionordered:

k=0.01/cmk=0.1/cmConductivity(referencedtowaterat77°Fk=0.01/cm0.04µS/cmto20µS/cmk=0.1/cm0.1µS/cmto200µS/cm

Temperature-22°to+320°F(-30°to+160°C

Pt100

TheConduMaxWisconnectedtothetransmitterusingthespecialmeasuringcableCYK71orCYK71-Ex

MeasuredvaluesCellconstantk

Measuringranges

TemperaturesensorCablespecifications

}

MeasuringcableCYK71/CYK71-Ex

ConduMaxWCLS12

Endress+Hauser

4Installation

Installationinstructions

Sensorsequippedwith1”NPT(orG1threadcanbemountedinaCLA751flowchamber(seeaccessoriesordirectlythreadedintoapipe.Whenmountingthe

sensor,makesurethatthemeasuringsurfacesarecompletelywettedbytheprocessfluidduringoperation.

Whenworkinginultrapurewater,ingressofairmustbeavoidedsincedissolvedair,particularlyCO2,mayincreasetheconductiivitybyupto3µS/cm.Theminimuminsertiondepthis2”(50mm.

Environment

IngressprotectionNEMA6(IP67

Refertoprocesstemperature

Ambienttemperature

Process

Processtemperature-22°to+320°F(-30°to+160°C

Maximum580psi(40barupto212°F(100°CwithoutCLA751flowchamberMaximum174psi(12barupto302°F(150°CwithCLA751flowchamber

Processpressure

Pressure/temperatureloadcurve

p[bar]

60

2030405030

–30

60

90

120

160

T[°C]

P-22

32

86

140194

248

320

87290435

580

725p[psi]T[°F]

PCLS12pressure/temperatureloadcurve

ConduMaxWCLS12

Endress+Hauser5

Mechanicalconstruction

Design,dimensions

CLSdesign

1Connectionhead2Nameplate

3Processconection

4Electrodes(coaxiallyarranged5Measuringcableconnection6SealingsurfaceaccordingtoDIN3852BI2,FormA

CLS12,dimensionsininches(mm

WeightApproximately3.1lb(1.4kgElectrodes

316TiSSProcessconnection316TiSS

ConnectionheadDie-castaluminumElectrodesealsEPDM,PEEK

1”NPTthreadorG1

1/2”NPTorPg16cablegland

Mateirals

ProcessconnectionCableconnection

Certificatesandapprovals

Hazardousapprovals

ATEXII1GEExiaIICT3/T4/T6

FMapproved,whenusedwithMyProCLM431andMycomSCLM153transmitters,contactEndress+Hauserforapproval.Withstatementoftheindividualcellconstant

Qualitycertificate

ConduMaxWCLS12OrderinginformationConduMaxWCLS12ConductivityMeasuringCellCLS121MeasuringrangeandcellconstantARange:

0.04-20µS/cm(k=0.01BRange:

0.1-200µS/cm(k=0.1Processconnection/material1DG1,316TiSS1K1”NPT,316TiSS1M3/4”NPT/PEScellshaft(onlyfixedcable1N1”NPT/PES(onlyfixedcable4AFlange,6hole,ANSI/316SS4BFlange,DN50,PN40,316SS4CFlange,ANSI,316SSCableconnection1Pg16cablegland51/2”NPTconduitentryTemperaturesensorAIntegratedPt100temperaturesensor1234234AccessoriesCLA751flowassemblyForinstallationofconductivitysensorswithG1threadonly.Inlet(bottomandoutlet(side,DN20withG1unionnuts.Material:

316TiSSMaximumtemperature:

320°F(160°CMaximumpressure:

174psi(12barOrdernumber:

500042016.30”(160Ø1.96”(50G1G14.72”(120Ø0.79”(20G1MeasuringcablesSpecialmeasuringcableCYK71fortwoelectrodeconductivitysensorswithintegratedtemperaturesensors,1low-noisecoaxialline,4auxiliarycoresat0.75mm2eachwithacommonshield,outerdiameter0.25”(7mm.Soldby1meterlengths,minimum15ft(5mOrdernumber:

5008533315ft(5mlengthOrdernumber:

5008828030ft(10mlengthOrdernumber:

50088281150ft(50mlengthOrdernumber:

50088284300ft(100mlengthOrdernumber:

50088285Specialmeasuringcable/extensioncableCYK71-Ex,forhazardousapplications,seeCYK71above.Specialcableisinbluesheath.Soldby1meterlengths,minimum15ft(5mOrdernumber:

500856736Endress+Hauser

ConduMaxWCLS12JunctionboxVBMjunctionboxforcableextension,with10screwterminals,NEMA4X(IP65Cableentry,1/2”NPTOrdernumber:

50003987Cableentry,Pg13.5Ordernumber:

51500177VBMjunctionboxforcableextensioninhazardousarea,with10high-impedancescrewterminals(blue,NEMA4X(IP65:

Ordernumber50003991CalibrationsolutionsCalibrationsolutionsareprecisionsolutionstoSRM(StandardReferenceMaterialofNISTforqualifiedcablibrationofconductivitymeasuringsystemsaccordingtoISO,accuracy±0.5%,withtemperaturetable.CLY11-A:

74µS/cm,referencetemperature77°F(25°C,500mlcontainerOrdernumber:

50081902CLY11-B:

149.6µS/cm,referencetemperature77°F(25°C,500mlcontainerOrdernumber:

50081903ConCalconductivitycalibrationsetisforultrapurewaterapplications,complete,factory-calibratedmeasuringsetwithcertificate,traceabletoSRMofNISTandDKD,comparativemeasurementinultrapurewaterapplicationsofupto10µS/cm.230VACunits,ordernumber:

50083777115VACunits,ordernumber:

50083778ConCalrecalibration,forfactoryrecalibrationandnewissueofcalibrationcertificate,traceabletoSRMofNISTandDKD,factorycalibrationprocedureaccordingtoASTMD-5391-93.Ordernumber:

51502486Forhightemperatureapplications,upto482°F(250°Candprocesspressuresupto580psi(40bar,requestConduMaxWCLS13conductivitysensor.RefertoTechnicalInformationTI083C.CalibrationsetRelatedproductsSupplementaldocumentationHazardousdocumentationConductivitysenosrsforapplicationinhazardousareasOrdernumber:

51512902MycomSCLM153transmitterTechnicalInformationLiquisysMCLM223/253transmitterTechnicalInformationMyProCLM431transmitterTechnicalInformationCLY11CalibrationsolutionsTechnicalInformationConCalCalibrationsetTechnicalInformationXA083C/07/a3TransmittersTI234C/24/aeTI193C/24/aeTI202C/24/aeTI162C/24/aeTI163C/24/aeCalibrationsolutions/CalibrationsetEndress+Hauser7

Forapplicationandselectionassistance,intheU.S.call888-ENDRESSFortotalsupportofyourinstalledbase,24hoursaday,intheU.S.call800-642-8737Visitusonourwebsite,UnitedStatesEndress+Hauser,Inc.2350EndressPlaceGreenwood,IN46143Phone:

(317535-7138888-ENDRESSFAX:

(317535-8498CanadaEndress+HauserCanadaLtd.1440Graham’sLaneUnit1,BurlingtonON,L7S1W3Phone:

(905681-9292800-668-3199FAX:

(905681-9444MexicoEndress+HauserPaseodelPedregalNo.610Col.JardinesdelPedregal01900,MexicoD.F.MexicoPhone:

(525568-2405FAX:

(525568-7459Endress+HauserThePowerofKnowHowTI082C/24/ae/07.03©2003Endress+Hauser,Inc.