DFMEA ReportWord下载.docx
《DFMEA ReportWord下载.docx》由会员分享,可在线阅读,更多相关《DFMEA ReportWord下载.docx(19页珍藏版)》请在冰豆网上搜索。
![DFMEA ReportWord下载.docx](https://file1.bdocx.com/fileroot1/2023-2/4/225b1732-fbde-499a-974c-19e6b7bd333b/225b1732-fbde-499a-974c-19e6b7bd333b1.gif)
3.dFMEATables4
4.Summary4
5.Conclusion5
(ToUpdateTable,ClickonTableArea,PressF9)
1.
GENERAL
1.1.Objective
DocumentandReporttheresultsoftheDesignFMEA(dFMEA)conductedonthedevicesubassembliesand/orcomponents.
1.2.Scope
TheUpgradeprojectisadesignchangeonWZ-50C6andWZ-50C6TMicroinfusionPumpSystems,whichwerereleasedintheyear2005andstillinsalesnow.
ThedFMEAisconductedonadeviceanditssubassembliesand/orcomponents.
1.3.Definitions
Term
dFMEA
DesignFailureModesandEffectsAnalysis
HA
HazardAnalysis–primaryproductriskanalysisdocument
1.4.CitedDocuments
Number
Revision
RD052
100
SmithsMedicalDesignFMEAProcedure
2.Methods(GeneralInstructions)
2.1.Team
IdentifythedFMEAteam.
Name
Title
2.2.Configuration
Thedevice,subassembliesand/orcomponentsthatwereanalyzedmustbeclearlyidentified.Partnumbersandrevisionsshallbenoted.
Restrictor
Footpad
Color
LCD
Label
Washer
Frontcase
Clamp
Switchsaddle
Buzzer
5ppmRisistor
Software
IFU
PressHandle
Cluchnetscrew
Leadscrew
Baseclutchscrew
Motor
Bearing
Battery
Transformer
Presstransducer
2.3.SystemEffectsListing
AlistingofallidentifiedSystemEffectsiscreatedduringtheanalysisinordertoserveasinputtothedeviceHazardAnalysis(HA).
Item
SystemEffect
Thepumpcannotstandstably
2.
Usernotsatisfied
3.
Losepartoffunctions
4.
Cannotcontinuetoprocess
5.
Shortcurcuit
6.
Difficulttoassemble
7.
Assemblyfails
8.
Safetyfunctionloses
9.
Incorrectalarm
10.
Usershutdowsthepumpaccidently
11.
Saftyissuewillbepossibletohappen
12.
Notuser-friendly
13.
Wrongoperation
14.
Thesyringecannotbefixed
15.
Thesyringcannotbeidentified
16.
Accuracyislow
17.
Reduecethelifetimeofcluchandleadscrews
18.
Thepumpcannotbeused
19.
Under-DeliveryofFluid
20.
Over-DeliveryofFluid
21.
NoDeliveryofFluid
22.
Leadandcluchscrewsarenotengagedwell
23.
ReducethepumpworktimewithoutACpower
2.4.ManufacturingControlsListing
AlistingofallidentifiedMfgControlsiscreatedduringtheanalysisinordertoserveasinputtothemanufacturingprocesses.NotethatrequirementsshouldbereferencedwhereassumedunderlyingfunctionalityispartoftheMfgControl(deviceselftestsetc.).
MfgControl
Reference
Addavisualinspectionstep
AdditionalIQC
PCBAFCT
AdditionalIQCandFQC
DetailedproductionWI
PCBICTandadditionaltraining
Supplierqualitycontrol
3.dFMEATables
SeeAppendixAforthedFMEATable(s).
4.Summary
[willbeaddedfinally]
5.Conclusion
AppendixA
dFMEATables
DFMEATables
Part
FailureMode
LocalEffect
PotentialCause
MfgControls
Comments
Screwlooses
Infusionlinecannotbefixedcorrectly
Noeffecttomainfunctions
ThetorqueofScrewdriverisnotenoughandbadqualityofscrew
Hookbroken
Infusionlinecannotbefixed
Lowqualityofhookandincorrectoperations
Footpadsarenotonthesameplane
Lowqualityoffootpadorwrongmaterial
Footpadsloose
Thescrewisnotfixed
Colorisnotconsistent
Rawmaterialcolorisnotright
Colorfades
Nodisplay
Cannotoperatebodyweightmode
LowqualityofLCD
brokenbyexternalforce
Nopower
Nodatacommunication
Nobacklight
Wrongcontents(includingcolorandword)
Usernotsatisfiedandnon-regulatorycompliance
Lowqualityoflabels
Stickinessisnotstrongenough
Cornerisbending
Lowqualityoflabelsandthelabelisnotadheredwell
Rearlabelandprintcodelabelarenotaligned
Codecannotbeidentified
Notgoodoperation
AdditionalFQC
Washerdoesn’tfunction
ThetorqueofScrewdriverisnotenough
Deformed
Pushrodisnotoperatedsmoothly
Difficulttoassemble,inaccuratepressure
Lowqualityoffrontcase
Dimensionofguidedblockisnotright
Clampisnotoperatedsmoothly
Mouldagingorpoormanufacturingprocesscontrol
AdditionalIQC?
Missinglocationofthebuzzerhole
PCBcannotbemounted
Jigfails
Changethemoldtomaketheholebysupplier
Dimensionofguidinggroove
AdditionalIQC
微动开关套
Screwholemislocation
Hardtomountmicroswitch
Thebuzzerismountedtoohigh
Impactfrontpanelstickprocess
Frontcasedeformed
Buzzerisnotinrightpositionwhenwelded
Buzzerdeosnotwork
Noalarmfunction
Safetyfunctionslose
Weldedwrongpin
24.
Componentdefect
25.
Positionisnotright
None
Volumeislower
26.
Lowqualityofbuzzersound
Lowqualityofbuzzer
27.
Sealedpaperisnottakenoff
ProductionWIandFQC
28.
Outoftolerence
Temperaturedrifttoomuch
Wrongalarm
Componentquality
29.
Pressingpowerkeytoshutdownthepumpevenduringinfusion
Thepumppowersoffdirectlynomatterifitisduringinfusion
Unclearspecification
30.
Syringesizechanedoesn’tcausethepumptostop
Thepumpcontinuestoinfuseevenifthesyringesizechanges
31.
Whileinfusingremovalofthesyringewillnotcauseanaudibleandvisiblealarm
Anaudibleandvisiblealarmdon’thappenevenifthesyringeisremovedwhileinfusing
32.
Whileinfusinganalarmwillnotcausethepumptostop
Thepumpcannotstopevenifanalarmhappenswhileinfusing
23.
Thepumpstartsaninfusionwhileanalarmisactive
Whenanalarmisactivethepumpstartsaninfusion
24
Thepumpdoesn’trestrictthevolumeoffluidduringasingleI/Vlinepurgeto2ml
25
DuringPOSTthepumpwill,withtheexceptionoftheon/offkey,testallkeystocheckwhetherakeyisstuckon
Somekeysarestuckon
26
Digitsafterthedecimalpointarenotonlycontrolledbysinglechevronkeys
Singleanddoublechevronkeyscontroldigitsafterthedecimalpoint
27
Notallofalarmsarebothaudibleandvisual
Somealarmsarenotbothaudibleandvisual
28
Keysarenotcheckediftheyarestuckoninbodyweightmode
Somekeysarepossibletobestuckon
29
Wrongversion
Difficultto/cannotunderstand
Carelessoperation
FQCandtraining
30
Ribweared
Holeofpresshandlebecomesbigger
Syringe